メインコンテンツへスキップ
Waters Japan

HDIで確認されるMaldiChrom処理エラー - WKB46116

Article number: 46116To English version

症状

  • HD Imaging(HDI)でデータを処理すると、エラーが確認されます
  • エラーの例:

clipboard_eecaaca274d6e313fc9f4862b254cd918.png

MaldiChrom v2.14 Feb-2016
Copyright 2008-2016 Micromass UK Ltd.
[PROCPARS:gpps.cpp(445)] Failed to find key 'Tolerance' in section [BACKSUB MALDICHROM PROCESS] in file C:\HDImaging\process\MaldichromParameter.ini.
[PROCPARS:gpps.cpp(445)] Failed to find key 'Flatten Edges’ in section [BACKSUB MALDICHROM PROCESS] in file C:\HDImaging\process\MaldichromParameter.ini.
[PROCPARS:gpps.cpp(445)] Failed to find key 'Type’ in section [BACKSUB MALDICHROM PROCESS] in file C:\HDImaging\process\MaldichromParameter.ini.
[PROCPARS:gpps.cpp(435)] Failed to find section [TOFACM MALDICHROM PROCESS] in file C:\HDImaging\process\MaldichromParameter.ini. [PROCPARS:gpps.cpp(445]] Failed to find key ’Centroiding Type’ in section [CENTER MALDICHROM PROCESS] in file C:\HDImaging\process \MaldichromParameter.ini.
[PROCPARS:gpps.cpp(445]] Failed to find key ’Use Areas’ in section [CENTER MALDICHROM PROCESS] in file C:\HDImaging\process\MaldichromParameter.ini.
0 lockmass adjustments to be applied
MSE Level 0

環境

  • HDIソフトウェア
  • SYNAPT G2-S/Si MALDI

原因

これらのメッセージは正常であり、.ini ファイルでパラメーターが指定されていない結果です。既定のパラメーターが使用されます。

解決策

なし。メッセージは正常です。

追加情報

id46116, MALDIG2SI, MALDIG2SI, MALDIMICRO, SYNG2SIHD, イメージング, ロックマス