HDI 1.4 で MaldiChrom 解析エラーが見られる - WKB46116
症状
- HD Imaging (HDI) v1.4 でデータを解析すると、エラーが観察される
- エラーの例:
MaldiChrom v2.14 Feb-2016
Copyright 2008-2016 Micromass UK Ltd.
[PROCPARS:gpps.cpp(445)] Failed to find key 'Tolerance' in section [BACKSUB MALDICHROM PROCESS] in file C:\HDImaging\process\MaldichromParameter.ini.
[PROCPARS:gpps.cpp(445)] Failed to find key 'Flatten Edges’ in section [BACKSUB MALDICHROM PROCESS] in file C:\HDImaging\process\MaldichromParameter.ini.
[PROCPARS:gpps.cpp(445)] Failed to find key 'Type’ in section [BACKSUB MALDICHROM PROCESS] in file C:\HDImaging\process\MaldichromParameter.ini.
[PROCPARS:gpps.cpp(435)] Failed to find section [TOFACM MALDICHROM PROCESS] in file C:\HDImaging\process\MaldichromParameter.ini. [PROCPARS:gpps.cpp(445]] Failed to find key ’Centroiding Type’ in section [CENTER MALDICHROM PROCESS] in file C:\HDImaging\process \MaldichromParameter.ini.
[PROCPARS:gpps.cpp(445]] Failed to find key ’Use Areas’ in section [CENTER MALDICHROM PROCESS] in file C:\HDImaging\process\MaldichromParameter.ini.
0 lockmass adjustments to be applied
MSE Level 0
環境
- HDI v1.4 ソフトウェア
- SYNAPT G2-S/Si MALDI
原因
これらのメッセージは正常であり、.ini ファイルでパラメーターが指定されていない結果です。既定のパラメーターが使用されます。
解決策
なし。メッセージは正常です。
追加情報
id46116, MALDIG2SI, MALDIG2SI, MALDIMICRO, SYNG2SIHD, イメージング, ロックマス